dong-ho-do-luu-luong-in-flow-f-110ci-bronkhorst-vietnam.png

Đồng hồ đo lưu lượng IN-FLOW F-110CI Bronkhorst Vietnam

IN-FLOW F-110CI

Industrial Style Thermal Mass Flow Meter for Gases

  • Lowest gas flow ranges
  • Compact design, small internal volume
  • High accuracy
  • Virtually pressure and temperature independent
  • Rugged, weatherproof housing (IP65, dust and waterproof)

Technical specifications

Measurement / control system
Flow range, based on N2
(intermediate ranges available)
min. 0,014…0,7 mln/min
max. 0,6…9 mln/min

Accuracy (incl. linearity)
(based on actual calibration)

±0,8% Rd plus ±0,2% FS for F-110CI-005;
±2% FS for F-110CI-002
Turndown 1:50 (in digital mode up to 1:187,5)
Multi fluid capability

storage of max. 8 calibration curves;
optional Multi Gas / Multi Range functionality up to 10 bar abs

Repeatability < 0,2% Rd

Response time

FS < 3 mln/min: 2 sec;
FS > 3 mln/min: 1…2 sec
Operating temperature -10…+70°C;
for ATEX cat. 3 and FM Class 1 Div 2 : 0…50°C
Temperature sensitivity zero: < 0,05% FS/°C; span: < 0,05% Rd/°C
Pressure sensitivity 0,1% Rd/bar typical N2; 0,01% Rd/bar typical H2
Leak integrity, outboard tested < 2 x 10-9 mbar l/s He
Attitude sensitivity max. error at 90° off horizontal 0,2% FS at 1 bar, typical N2
Warm-up time 30 min. for optimum accuracy
2 min for accuracy ± 2% FS
Mechanical parts
Material (wetted parts) stainless steel 316L or comparable
Pressure rating 100 bar abs
Process connections compression type or face seal couplings
Seals standard: Viton®; options: EPDM, Kalrez® (FFKM)
Weight 0,8 kg
Ingress protection (housing) IP65
Electrical properties
Power supply +15…24 Vdc
Max. power consumption Meter Supply
15 V
24 V
at voltage I/O
95 mA
65 mA
at current I/O
125 mA
85 mA
Max. power consumption Controller Supply
15 V
24 V
at voltage I/O
290 mA
200 mA
at current I/O
320 mA
215 mA
PROFIBUS DP add 53 mA (15 V supply) or 30 mA (24 V supply)
PROFINET add 77 mA (15 V supply) or 48 mA (24 V supply)
DeviceNet™ add 48 mA (24 V supply)
Analog output/command 0...5 (10) Vdc or 0 (4)…20 mA (sourcing output)
Digital communication standard: RS232;
options: PROFIBUS DP, PROFINET, DeviceNet™, Modbus RTU or ASCII, FLOW-BUS
Electrical connection
Analog/RS232 8 DIN (male);
PROFIBUS DP bus: 5-pin M12 (female);
power: 8 DIN (male);
PROFINET bus: 2 x 5-pin M12 (female) (in/out);
power: 8 DIN (male);
Devicenet™ 5-pin M12 (male);
FLOW-BUS/Modbus-RTU/ASCII 5-pin M12 (male)